OmniaLabDS
The reliable one.
For H2O pure types II + CLRW (CLSI) + DIN EN 285 + EN ISO 15883.
When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.
Features
- OptiFill dispenser is standard
- 100 l tank with quality recirculation and pressure outlet
- Tank volume display in percent
- Tank volume can be modularly increased
- Simple, cost-effective filter replacement
- Leakage sensor is standard
- Emergency supply (optional)
- Degassing unit (optional)
specifications
OmniaLabDS20 | OmniaLabDS40 | OmniaLabDS60 | OmniaLabDS80 | ||
---|---|---|---|---|---|
Type II + CLRW (CLSI) DIN EN 285 + ISO EN 15883 | |||||
Pure Water Performance(l/h at 15 °C) |
20 | 40 | 60 | 80 | |
Conductivity(μS/cm) | 0.1 – 1.0 | 0.1 – 1.0 | 0.1 – 1.0 | 0.1 – 1.0 | |
Resistance(M Ω x cm) | 10 – 1 | 10 – 1 | 10 – 1 | 10 – 1 | |
Dispensing Performance Dispenser(l/min.) | up to 2 | up to 2 | up to 2 | up to 2 | |
Pure Water Tank Pressurized Outlet | optional | optional | optional | optional | |
Particles*(/ml) | < 1 | < 1 | < 1 | < 1 | |
Bacteria* (CFU/ml) | < 0.1 | < 0.1 | < 0.1 | < 0.1 | |
Feedwater Requirements |
|||||
Softened or hardness-stabilized water according to DIN 2000 | |||||
Feedwater Temperature(°C) | + 2 up to 35 | + 2 up to 35 | + 2 up to 35 | + 2 up to 35 | |
Manganese and Iron Content(mg/l) | < 0.05 | < 0.05 | < 0.05 | < 0.05 | |
Free Chlorine Content(mg/l) | < 0.1 | < 0.1 | < 0.1 | < 0.1 | |
Silt Density Index(SDI) | max. 3 | max. 3 | max. 3 | max. 3 | |
Technical Data |
|||||
Operating Pressure(bar) | 2 – 6 | 2 – 6 | 2 – 6 | 2 – 6 | |
Supply Voltage(Volt/Hz) | 90-240/50-60 | 90-240/50-60 | 90-240/50-60 | 90-240/50-60 | |
Connected Load(kW) | 0.1 | 0.1 | 0.1 | 0.1 | |
Connector Size | R 3/4″ | R 3/4″ | R 3/4″ | R 3/4″ | |
Ambient Temperature(°C) | + 2 up to 35 | + 2 up to 35 | + 2 up to 35 | + 2 up to 35 | |
Dimensions Tower**(W x H x D mm) |
511 x 1520 x 575 | 511 x 1520 x 575 | 511 x 1520 x 575 | 511 x 1520 x 575 | |
Dimensions Base Cabinet Tank(mm) |
511 x 800 x 575 | 511 x 800 x 575 | 511 x 800 x 575 | 511 x 800 x 575 | |
Weight(kg – without polishing cartridge) | 39 | 39 | 40 | 40 |
* With sterilizing filter 0.2 μm.
** With OptiFill Dispenser.
Manufacturers
show DATA — DO NOT CHANGE ID–
Industries
show DATA — DO NOT CHANGE ID–
Methods
show DATA — DO NOT CHANGE ID–
Spares and accessories for OmniaLabDS
Pre-treatment unit OmniaLab – 10“ (19200020)
UV tank disinfection unit 254 nm (19200050)
UV Flow through disinfection 254 nm (19200051)
Replacement/second polishing cartridge type DS 2800 RV (12280050)
Emergency supply (19200040)
Degassing unit (19200041)
External pressure booster pump SC 3000 (16561201