HR4PRO-UV-VIS-ES
HR spectrometers are small-footprint, high-resolution spectrometers for applications such as wavelength characterization of lasers and LEDs, monitoring of gases and monochromatic light sources, and determination of elemental atomic emission lines. Depending on spectrometer configuration, optical resolution <0.1 nm (FWHM) is possible.
New HR4Pro preconfigured spectrometers are compact, high-resolution options distinguished by low stray light performance and great thermal stability for applications in lab, field and process environments. With a robust optical bench design and optimized components, the HR4Pro provides up to 10x improvement in thermal stability compared with similar small-bench spectrometers. Models are available for UV-Vis (200-875 nm), Vis-NIR (350-1025 nm) and extended-range (200-1100 nm) measurements.
Features
- High resolution performance
- New preconfigured options
- Affordable
- Convenient integration
- Flexible triggering functions
- High speed acquisition
specifications
Spectroscopy: |
|
---|---|
Wavelength Range | 200nm – 875nm |
Optical Resolution |
~0.66nm FWHM |
Integration Time | 3.8ms – 10s |
Corrected linearity | 0.5% NL |
Dark Noise | 6 counts RMS |
Dynamic Range | 2500 |
Input Fiber Connector | SMA 905 |
Signal to Noise Ratio | 300:1 (full signal) |
Thermal Stability |
0.03 pixels/°C |
Detector: |
|
Grating | GRATING_#H48 |
Pixels |
3648 |
Detector | linear silicon CCD array |
Entrance slit | 10 µm |
Order Sorting | OTF4 |
Detector Collection Lens | Yes |
Electronics: |
|
A/D Resolution |
14-bit |
Power Requirement | 5VDC, 500ma |
Communications: |
|
Wired |
USB 2.0 |
Environmental: |
|
Compliance | RoHS, CE |
Humidity | 0-90% non-condensing |
Material | Aluminum |
Operating Temperature | -10◦to +50◦C |
Storage Temperature | -30° to 70° C |
Physical: |
|
Dimensions | 148.6mm X 104.8mm X 45.1mm |
Weight | 570g |
Manufacturers
show DATA — DO NOT CHANGE ID–
Industries
show DATA — DO NOT CHANGE ID–
Methods
show DATA — DO NOT CHANGE ID–
For More Information Please Click Here
Spares and accessories for HR4PRO-UV-VIS-ES
HR4-CBL-DB15
Connecting the Flame and X-platform spectrometers is easy with our range of cables and connectors. These spectrometers have a connector called the DD4 connector. It has a low profile, high density 40 pin connector type, similar to those used on many modern consumer electronic devices, such as tablets and smartphones. This lets us pack a lot into a small space!