QEPRO-ABS
The QE Pro is a high sensitivity spectrometer with low stray light performance. It is ideal for a wide range of low light level applications such as fluorescence, DNA sequencing and Raman analysis. Custom and preconfigured absorbance and fluorescence options are available.
The sensitivity and large dynamic range of the QE Pro combine to make it the highest-performance miniature spectrometer in its class. Absorption, reflectance, fluorescence and Raman measurements all benefit from the superior performance of the QE Pro. For advanced measurements, spectral buffering ensures data integrity at high collection rates; advanced optoelectronics and thermoelectric cooling combine to provide thermal stability over long-term measurements.
Features
- Sensitive
- Powerful
- Stable
- Configurable
- Buffering
- Preconfigured for absorbance
specifications
Communications: |
|
---|---|
Wired | USB 2.0, 480 Mbps (USB 1.1 compatible); RS-232 (5-wire) |
Detector: |
|
Detector | Back-thinned, TE Cooled, 1024 x 58 element CCD array |
Entrance slit | 10 µm |
Grating | GRATING_#HC1-QE |
Pixels | 1024 |
Detector Collection Lens | No |
Order Sorting | OFLV-QE-200 |
Quantum Efficiency | 90% (peak) |
Electronics: |
|
Power Requirement | Supply voltage: 4.5 – 5.5 V |
Strobe Functions | continuous and single strobe |
Environmental: |
|
Humidity | ≤ 90% noncondensing |
Operating Temperature | 0 °C – 50 °C |
Physical: |
|
Dimensions | 182 mm (7.17 in.) x 110 mm (4.33 in.) x 47 mm (1.85 in.) |
Weight | 1.15 kg (2.6 lbs.); power supply: 0.45 kg (1 lb.) |
Spectroscopy: |
|
Integration Time | 8ms – 3600s |
Optical Resolution | 1.70 nm FWHM (typical) |
Wave Length Range | 200nm – 950nm |
Dynamic Range | ~85,000:1 |
Input Fiber Connector | SMA 905 |
Onboard Memory | 15,000 spectra |
Signal to Noise Ratio | 1000:1 (single acquisition) |
Stray light | <0.08% at 600 nm; 0.4% at 435 nm |
Manufacturers
show DATA — DO NOT CHANGE ID–
Industries
show DATA — DO NOT CHANGE ID–
Methods
show DATA — DO NOT CHANGE ID–
* For More Information Please Click Here
Spares and accessories for QEPRO-ABS
HR4-CBL-DB15
Connecting the Flame and X-platform spectrometers is easy with our range of cables and connectors. These spectrometers have a connector called the DD4 connector. It has a low profile, high density 40 pin connector type, similar to those used on many modern consumer electronic devices, such as tablets and smartphones. This lets us pack a lot into a small space!