.
SR-2XR250
500 Lines Blazed at 250 nm
SR-2XR250
Specifications
Entrance Slit | SR-2XR250 | SR-2XR250 | SR-2XR250 | SR-2XR250 | SR-2XR250 | SR-2XR250 |
Optical and Spectroscopic | ||||||
Wavelength Range (configuration-dependent) |
220-1050 | |||||
Resolution (FWHM) (configuration-dependent) |
1.44 | 1.54 | 2.02 | 3.12 | 5.76 | 11.52 |
Integration Time | 1 μs – 2 s | |||||
OSF | Y | |||||
Dynamic Range for single acquisition1 | 3400:1 | |||||
Signal-to-Noise (maximum SNR per 1 second w/High Speed Averaging Mode) | 25800:1 | |||||
Signal to Noise
(Single scan @ 10 ms) |
380:1 | |||||
Scan rate (type)² | 630 Hz | |||||
Thermal Wavelength Drift | 0.02 nm/°C | |||||
Triggering | Software, External rising edge, External level | |||||
Grating | Configurable | |||||
Detector | ||||||
Active Pixels | 2098 | |||||
Corrected Linearity | 99.7% | |||||
Electrical | ||||||
Connectors | USB Type-C, SMA, 16 pin Samtec TFM, RS-232 | |||||
Power-up Time | 3 s | |||||
Power Requirements | 5VDC | |||||
Mechanical | ||||||
Physical Dimensions | 88.1 x 63.5 x 31.45 mm | |||||
Weight | 275 g | |||||
Input Fiber Connector | SMA 905 | |||||
Environmental | ||||||
Temperature (operation) | 0 °C – 55 °C | |||||
Temperature (storage) | -30 °C to 70 °C | |||||
Humidity | <85% | |||||
Operation | For indoor use | |||||
Operating Altitude (maximum) | 2000 m | |||||
Overvoltage Category | 1 | |||||
Pollution Degree | 2 |
Manufacturers
show DATA — DO NOT CHANGE ID–