Wavelength Calibration Sources
Spectrometer wavelength calibration sources are available for convenient, reliable spectrometer…
Show Radiometrically calibrated light sources from Ocean Insight are used to calibrate the absolute spectral response of a spectrometer system. Using these light sources and our software, you can determine absolute intensity values across UV, Visible and NIR wavelengths.
Radiometrically calibrated light sources are carefully characterized to deliver a known quantity of light with very low uncertainty. Each source is measured to NIST-traceable standards, creating an unbroken chain of traceability and providing you with the highest quality and most reliable data.
Subject to change depending on model type
Engineering: |
|
---|---|
Nominal Bulb Power | 25 W (deuterium), 20 W (tungsten halogen), 5 W |
Source | Deuterium & Tungsten Halogen, Tungsten halogen |
Wavelength Range | 210nm – 1.1μm |
Calibrated For | Absolute irradiance (µW/cm2/nm), Absolute irradiance (µW/cm2/nm) |
Calibration Lifetime | 50 hours |
Drift (Voltage) | <0.01% per hour, <0.3% per hour |
Minimum Uncertainty Level | 3%, 3.7% at 700nm, 4.0% at 700nm, 7% |
Stability (Current) | <5 x 10-6 peak-to-peak (0.1-10.0 Hz), 0.5% peak-to-peak |
Warm-up Time | 15 minutes, 40 minutes (deuterium), 40 minutes (deuterium), 20 minutes (tungsten halogen) |
Environmental: |
|
Safety & Regulatory | CE, ROHS, WEEE |
Operating Humidity Specification | 5-95% without condensation at 40 °C, 5-95% without condensation at 40°C |
Operating Temperature | 5 °C – 35 °C |
Physical: |
|
Fiber Connector | Integrating sphere, SMA 905; CC-3, SMA 905; CC-3-UV-S or 6.35 mm barrel for cosine corrector |
Shutter | Yes |
Dimensions | 14 x 6 x 6 cm, 15 x 13.5 x 28.5 cm |
Power Consumption | 1.2 A @ 12 VDC, Approximately 78VA |
Power Requirements | 85-264 V, 50/60 Hz |
Weight | 0.5 kg, 6 kg |