Vertex.5A

The Vertex.5A is a versatile entry-level potentiostat which has ±5A / ±10V compliance and optional integrated impedance analyser up to 1MHz.

The Vertex.5A has a peripheral port to allow external interfacing and compatibility with several Ivium modules (see expandability).

The Vertex can optionally be expanded with a True Linear Scan module, Bipotentiostat, and/or FRA/EIS capability. Next to that the Vertex is compatible with other Ivium options and modules, such as the HiMUX and CIM, and several Ivium boosters for higher power requirements.

Do NOT CHANGE OR REMOVE THIS ELEMENT

Features

  • Scan range: ±5A/±10V
  • Optional FRA/EIS: 10µHz to 1MHz
  • Optional True Linear Scan generator
  • Optional Bipotentiostat

Specifications

System performance
Number of channels in instrument
1
Current compliance ±5A
Maximum compliance voltage ±10V
Maximum applied voltage ±10V
Electrode connections 4; WE, CE, RE, S (and GND) with 4mm banana plugs
Potentiostat bandwidth >500kHz
Stability settings High Speed, Standard and High Stability
Programmable response filter 1MHz, 100kHz, 10kHz, 1kHz, 10Hz
Signal acquisition Dual channel 18bit ADC, 300,000 samples/s
Potentiostat
Applied potential range ±10V, 0.08mV res.
Applied potential accuracy 0.2%, or 2mV
Current ranges ±100pA to ±10A
Measured current resolution 0.003% of current range, minimum 3fA
Measured current accuracy 20pA + 0.025% of FSR
Galvanostat
Applied current resolution 0.008% of applied current range
Applied current accuracy 0.2%
Galvanostatic current ranges ±10nA to ±10A
Potential ranges ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
Measured potential resolution 0.0008% of potential range, minimum 7nV
Measured potential accuracy 0.2% or 2mV
Impedance Analyser
Frequency range 10µHz to 1MHz
Amplitude 0.15mV to 2.0V, or 0.03% to 100% of current range
DC offset 16bit DC offset subtraction and 2 DC-decoupling filters
Electrometer
Input impedance >1000Gohm //<10pF
Input bias current <20pA
Bandwidth >5MHz
Special functions
Ohmic drop compensation 2V/current range
Floating operation User selectable
Peripheral Connections
Analog in/out 2/1 (±10V, 16bit resolution, bandwidth 40kHz)
Digital input/output 1 input (0-5V) / 3 output (0-3.3V)
I-out/E-out Analog monitor for cell current and potential
AC-out ±0.5V sinewave, 10µHz-1MHz with var. att.
Channel-X input, Channel-Y input
±4V: to record impedance from peripheral devices
Bipotentiostat (optional)
Current compliance (BiStat) ±35mA
Applied potential offset ±2V with 0.06mV resolution wrt. RE or WE
Environment
Power requirements 100-240V, 50-60Hz, 2A
Interfacing USB
Size (w x d x h) 15 x 36 x 4.5cm
Weight 2.5kg
PC requirements Windows 8/10/11, with free USB port