PRODUCT

CompactStat.h standard

The CompactStat.h is a truly portable yet powerful potentiostat which has ±30mA / ±10V compliance and integrated impedance analyser up to 3MHz.

Multiple analog and digital input & output ports are available for monitoring and control of peripheral equipment. This functionality is fully integrated in the software.

The CompactStat.h can be expanded with a wide variety of options and modules, including a Bipotentiostat, True Linear Scan generator, various power boosters and multiplexers.

Do NOT CHANGE OR REMOVE THIS ELEMENT

Features

  • Scan range: ±30mA/±10V
  • FRA/EIS: 10µHz – 3MHz

Specifications

System performance
Current compliance ±30mA
Maximum compliance voltage ±10V
Maximum applied voltage ±10V
Electrode connections 4; WE, CE, RE, S (and GND) with 4mm banana plugs
Potentiostat bandwidth >3MHz
Stability settings High Speed, Standard and High Stability
Programmable response filter 1MHz, 100kHz, 10kHz, 1kHz, 10Hz
Signal acquisition
Dual channel 24bit ADC, 100,000 samples/s
Potentiostat
Applied potential range ±4V, 0.01mV res. (20bit)/±10V, 0.02mV res.
Applied potential accuracy 0.2% or 1mV
Current ranges ±1pA to ±1A in 14 decades
Measured current resolution 0.00001% of current range, minimum 0.6aA
Measured current accuracy 10pA + 0.025% of FSR
Galvanostat
Applied current resolution 0.00013% of applied current range
Applied current accuracy 0.2%
Galvanostatic current ranges ±100pA to ±1A
Potential ranges ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V
Measured potential resolution 0.00001% of potential range, minimum 0.05nV
Measured potential accuracy 0.2% or 1mV
Impedance Analyser
Frequency range 10µHz to 3MHz
Amplitude 0.015mV to 1.0V, or 0.03% to 100% of current range
DC offset 16bit DC offset subtraction and 2 DC-decoupling filters
Dynamic range 0.05nV to 10V, and 0.2aA to 30mA
Electrometer
Input impedance >1000Gohm //<8pF
Input bias current <10pA
Bandwidth >16MHz
Special functions
Ohmic drop compensation 2V/current range
Floating operation User selectable
Peripheral connections
Analog in/out 8/2 (0 to +4V, 16bit resolution)
Digital input/output 2 input (0-5V) / 3 output (0-3.3V)
I-out/E-out Analog monitor for cell current and potential
AC-out ±0.5V sinewave 10µHz-3MHz with variable attenuation
Channel-X input, Channel-Y input ±4V: to record impedance from peripheral devices
Environment
Power requirements In USB powered mode: standard 5V, 500mA, External adapter: 100-240V, 50-60Hz, 350mA
Interfacing USB
Size (w x d x h) 12 x 26 x 2.5cm
Weight 0.6kg
PC requirements Windows 8/10/11, with free USB port