Vertex.1A
The Vertex.1A is a versatile entry-level potentiostat which has ±1A / ±10V compliance and optional integrated impedance analyser up to 1MHz.
The Vertex.1A has a peripheral port to allow external interfacing and compatibility with several Ivium modules (see expandability).
The Vertex.1A can optionally be expanded with a True Linear Scan module, Bipotentiostat and/or FRA/EIS capability. Next to that the Vertex is compatible with other Ivium options and modules, such as the HiMUX and CIM, and several Ivium boosters for higher power requirements.

Features
- Scan range: ±1A/±10V
- Optional FRA/EIS: 10µHz to 1MHz
- Optional True Linear Scan generator
- Optional Bipotentiostat
Specifications
System performance |
|
|---|---|
|
Number of channels in instrument
|
1 |
| Current compliance | ±1A |
| Maximum compliance voltage | ±10V |
| Maximum applied voltage | ±10V |
| Electrode connections | 4; WE, CE, RE, S (and GND) with 4mm banana plugs |
| Potentiostat bandwidth | >500kHz |
| Stability settings | High Speed, Standard and High Stability |
| Programmable response filter | 1MHz, 100kHz, 10kHz, 1kHz, 10Hz |
| Signal acquisition | Dual channel 18bit ADC, 300,000 samples/s |
Potentiostat |
|
| Applied potential range | ±10V, 0.08mV res. |
| Applied potential accuracy | 0.2%, or 2mV |
| Current ranges | ±100pA to ±1A |
| Measured current resolution | 0.003% of current range, minimum 3fA |
| Measured current accuracy | 20pA + 0.025% of FSR |
Galvanostat |
|
| Applied current resolution | 0.008% of applied current range |
| Applied current accuracy | 0.2% |
| Galvanostatic current ranges | ±10nA to ±1A |
| Potential ranges | ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V |
| Measured potential resolution | 0.0008% of potential range, minimum 7nV |
| Measured potential accuracy | 0.2% or 2mV |
Impedance Analyser |
|
| Frequency range | 10µHz to 1MHz |
| Amplitude | 0.15mV to 2.0V, or 0.03% to 100% of current range |
| DC offset | 16bit DC offset subtraction and 2 DC-decoupling filters |
Electrometer |
|
| Input impedance | >1000Gohm //<10pF |
| Input bias current | <20pA |
| Bandwidth | >5MHz |
Special functions |
|
| Ohmic drop compensation | 2V/current range |
| Floating operation | User selectable |
| Peripheral Connections | |
| Analog in/out | 2/1 (±10V, 16bit resolution, bandwidth 40kHz) |
| Digital input/output | 1 input (0-5V) / 3 output (0-3.3V) |
| I-out/E-out | Analog monitor for cell current and potential |
| AC-out | ±0.5V sinewave, 10µHz-1MHz with var. att. |
| Channel-X input, Channel-Y input |
±4V: to record impedance from peripheral devices
|
| Bipotentiostat (optional) | |
| Current compliance (BiStat) | ±35mA |
| Applied potential offset |
±2V with 0.06mV resolution wrt. RE or WE
|
Environment |
|
| Power requirements | 100-240V, 50-60Hz, 2A |
| Interfacing | USB |
| Size (w x d x h) | 13 x 27 x 4cm |
| Weight | 1.5kg |
| PC requirements | Windows 8/10/11, with free USB port |