Seta Multiflash Cleveland Flash Point Module

The automatic Cleveland module conforms precisely to national and international Cleveland open cup flash point test methods. It comprises a sample cup, temperature probe, flash/fire detector, and a DIPS pod containing the sweeping mechanism with co-axial gas/electric ignitor.
The Cleveland module must be used with the Multiflash Automatic Flash Point – Universal Base Unit (34000-0). The base unit recognises that the Cleveland module is connected and instantaneously sets up standard test parameters and calibration data.

Do NOT CHANGE OR REMOVE THIS ELEMENT

Features

  • Fully self-contained flash/fire point test module
  • Ambient to 400 °C
  • Gas or electric ignition
  • Automatic heating control, flash and fire detection
  • Automatic test results & end of test audible warning
  • Automatic snuffer
  • Automatic barometric correction

Specifications

Temperature Range Ambient to 400 °C
Bath N/A
Heating Hotplate
Post Test Cooling Fan assisted or Forced Air (Requires 34008-0 Cooling Boost Module option)
Post Test Cool-Down Time (90°C TO 50°C) 9 minutes (Fan) 6 minutes (Forced Air)
Ignitor Quick-fit coaxial gas jet / electric hot wire, user selectable (Electric hot wire pilot light for gas jet)
Temperature Probe 30 rpm
Gas Supply (Gas Jet Ignitor) Class A Platinum Resistance Thermometer, °C or °F
Flash/ Fire Detection Ionisation Ring
Gas Supply (Gas Jet Ignitor) Laboratory gas, 3 kPa (0.44 psi) maximum pressure
Air Supply (Forced Air Cooling) Dry compressed air, 670 kPa (100 psi) maximum pressure
Voltage 110/120 V or 220/240 V, 50/60 Hz, (switchable)
Power 1 kW
Size (HXWXD) 330 x 470 x 420 mm
Weight 12 kg