PalmSens4
PalmSens flagship instrument, the PalmSens4, is a USB and battery powered Potentiostat, Galvanostat, and optional a Frequency Response Analyser (FRA) for Electrochemical Impedance Spectroscopy (EIS). The PalmSens4 has a large potential range (-5V to 5V or -10V to 10V) and current range (100 pA to 10 mA) with a high resolution and low noise. The economical PalmSens4 is a complete laboratory instrument but its compact and rugged design makes it also ideal for field work. Connecting via Bluetooth guarantees a perfectly floating measurement.

Features
- ±5 V or ±10 V potential range
 - EIS/FRA with maximum frequency of 100 kHz or 1 MHz
 - optional Bipotentiostat module for second WE
 - optional iR-Compensation
 - Rugged carrying case
 - High quality, double shielded cell cable with 2 mm banana connectors for Working, Counter, Reference electrode and Ground
 - Crocodile clips
 - Dummy cell
 - USB cable
 - Manual and Quick Start document
 - PSTrace software for Windows
 
Specifications
General | 
|
|---|---|
| dc-potential range | ±10 V (or ±5 V) * | 
| compliance voltage | ±10 V | 
| maximum current | ±30 mA (typical) | 
| max. acquisition rate | 150,000 data points/s | 
Potentiostat (controlled potential mode) | 
|
| applied dc-potential resolution | 76.3 µV (18-bit) | 
| applied potential accuracy | ≤0.1% ±1 mV offset | 
| current ranges | 100 pA to 10 mA (9 ranges) | 
| current accuracy | ≤0.1 % (at Full Scale Range) | 
| measured current resolution | 0.005 % of current range (18-bit, 5 fA on 100 pA range) 0.0025% of 10 mA range  | 
Galvanostat (controlled current mode) | 
|
| current ranges | 1 nA to 10 mA (8 ranges) | 
| applied dc-current range | ±6 times applied current range | 
| applied dc-current resolution | 0.0076% of applied current range (<10 mA) 0.0038% of 10 mA range  | 
| measured dc-potential resolution | 78 μV at ±10 V (gain 1, 18-bit) 7.8 μV at ±1 V (gain 10) 0.78 μV at ±0.1 V (gain 100)  | 
| measured dc-potential accuracy | ≤ 0.05% or ±1 mV (for |E| < ±9 V) ≤ 0.2% (for |E| ≥ ±9 V)  | 
FRA / EIS (impedance measurements) | 
|
| frequency range | 10 μHz to 1 MHz (or 10 μHz to 100 kHz) * | 
| ac-amplitude range | 1 mV to 0.25 V rms, or 0.7 V p-p | 
GEIS (galvanostatic impedance measurements) | 
|
| frequency range | 10 μHz to 100 kHz (or 10 μHz to 100 kHz) | 
| ac-amplitude range | 0.001 x CR to 0.4 x CR (<10 mA) 0.001 x CR to 0.2 x CR (10 mA) (CR = current range)  | 
Electrometer | 
|
| electrometer amplifier input | > 1 TΩ // 10 pF | 
| bandwidth | 1 MHz | 
Other | 
|
| housing | aluminium with rubber sleeve: 15.7 x 9.7 x 3.5 cm3 | 
| weight | +/- 500 g | 
| temperature range | 0 ºC to + 50 ºC | 
| power supply | USB or internal LiPo battery | 
| communication | USB and Bluetooth (Dual Mode) | 
| battery time | > 16 hours idle time > 4 hours with cell on at max. current extendable by means of power bank  | 
| internal storage space | 8 GB or +/- 800000 measurements incl. method info (assuming 200 data points per measurement)  | 
Auxiliary port (D-Sub 15) | 
|
| analog input | ±10 V, 18-bit | 
| analog output | 0-10 V, 12 bit (1 kOhm output impedance) | 
| 4 digital outputs | 0-5 V | 
| 1 digital input | 0-5 V | 
| i-out and E-out | raw output of current and potential E-out ±10 V (1 kOhm output impedance) i-out ±6 V (1 kOhm output impedance)  | 
| power | 5 V output (max. 150 mA) | 
* Depending on PalmSens4 hardware configuration
Techniques
Voltammetric techniques
- Linear Sweep Voltammetry (LSV)
 - Cyclic Voltammetry (CV)
 - Fast Cyclic Voltammetry (FCV)
 - AC Voltammetry (ACV)
 
Pulsed techniques
- Differential Pulse Voltammetry (DPV)
 - Square Wave Voltammetry (SWV)
 - Normal Pulse Voltammetry (NPV)
 
Galvanostatic techniques
- Linear Sweep Potentiometry (LSP)
 - Chronopotentiometry (CP)
 - MultiStep Potentiometry (MP)
 - Open Circuit Potentiometry (OCP)
 - Stripping Chronopotentiometry (SCP or PSA)
 
Amperometric techniques
- Chronoamperometry (CA)
 - Zero Resistance Amperometry (ZRA)
 - Chronocoulometry (CC)
 - MultiStep Amperometry (MA)
 - Fast Amperometry (FAM)
 - Pulsed Amperometric Detection (PAD)
 - Multiple-Pulse Amperometric Detection (MPAD)
 
Other
- Mixed Mode (MM)
 
Potentiostatic/Galvanostatic Impedance spectroscopy (EIS/GEIS)
- Potential scan or current scan
 - Fixed potential or fixed current
 - Time scan
 