Printer paper, pack of 20.
AvCount Lite Particle Counter – SA1800-2
The AvCount Lite portable automatic particle counter provides reliable results for determining the size and distribution of particles in light and middle distillate fuels, including aviation fuel and kerosene, biodiesel, low viscosity oils and hydraulic oils.
The particle counters simple operation makes testing bottle samples or on-line quick and easy, with results in under 3 minutes.
The AvCount Lite displays the cumulative, particles/ml and ISO 4406 cleanliness codes. Up to 600 measurements can be saved to the internal memory, downloaded to a PC/laptop or printed via the internal printer.
Supplied with: sample bottles, sample probe, waste pipe, universal voltage power supply, USB cable, calibration certificate and a USB memory stick containing handbooks and Partikel software for basic test method programming and data download. Learn more about
AvCount Lite Particle Counter – SA1800-2
Features
- ISO 11171 calibration
- Cumulative Particles/ml
- ISO 4406 Cleanliness Codes
- Portable, compact instrument
- Online/offline (bottle sample)
- Stand-alone or PC controlled
- Programmable via PC
- Integrated printer
- Battery power optional
specifications
PARTICLE SIZE RANGE | ISO 11171: 4 µm(c) to >70 µm© (calibration for larger sizes available on request) |
---|---|
NUMBER OF MEASURING CHANNELS | 6 size channels displayed on instrument, up to 16 programmable via PC 4 µm(c), 6 µm(c), 10 µm(c), 14 µm(c), 21 µm(c), 25 µm(c), 30 µm(c), 38 µm(c), 70 µm(c) |
COUNTS PER MEASUREMENT (MAX) | 60,000 per ml |
SAMPLE VISCOSITY (MAX) | 64 mm2/s (using internal pump), 200 mm2/s (pressure fed @ 3 BarG) (SA1950-0 Sample Delivery System is available as an accessory) |
SAMPLE TEMPERATURE RANGE | Ambient 0 to 70 °C |
RESULTS | 900 result memory Date/time indexed (last 20 tests can be viewed on screen) Download to PC/laptop port or ProTrend software and print via internal printer |
SAMPLE FLOW RATE | 30 ml/min ±5 ml/min |
TOTAL SAMPLE VOLUME USED | 80 ml for ASTM D7619 & IP 565, from 20 ml for other methods (includes flush cycles) |
VOLTAGE | 12 Vdc 3A with printer on, 1A with printer off Charger/mains adaptor 100 to 240 Vac, 50/60Hz input, 12 Vdc 2.5A output Optional 12 Vdc battery and charger (SA1816-0) |
SIZE (HXWXD) | 250 x 145 x 350 mm |
WEIGHT | 7.5 kg |
CCCN CODE | TARIFF 90278090 |
Manufacturers
show DATA — DO NOT CHANGE ID–
Industries
show DATA — DO NOT CHANGE ID–
Methods
show DATA — DO NOT CHANGE ID–
Spares and accessories for SA1800-2
Printer Paper (pack of 20) – 81002-001
AvCount Replacement Sample Probe – SA1000-001
AvCount Exit Tubing and Connector – SA1000-004
Exit tubing and connector with FFKM O-ring for use with AvCount2 (SA1000-2).
Ultrasonic Bath to ISO 11171 – 99320-3
AvCount 250 ml Verification Material – SA1006-0