Seta Multi Filtration Tester (MFT) – 91600-4
The Seta Multi Filtration Tester (MFT) is a fully automated instrument, designed to test the Filter Blocking Tendency (FBT) of diesel, biodiesel (B100 & B5/7/20/30), gas oil, gas turbine fuel and kerosene.
The MFT features a colour touch screen and clear, simple user interface which guides the operator through the menu system and simple operating sequence. Results are stored internally and can be viewed on the screen or saved directly to a USB memory stick or LIMS for further data management and analysis. Alternatively, a printer can be connected using the RS232 serial port.
Designed to be compact and lightweight, the MFT’s small footprint makes it ideal for laboratories with limited bench space.
FBT analysis is achieved by measuring the pressure differential across a disposable filter. Fuels which have a high FBT could potentially block filters in the distribution network or during use in a vehicle or power plant.
MFT is also used to determine the FBT for BX fuels containing bio components, such as FAME (biofuel), which have been ‘cold soaked’ at a temperature just above zero Celsius. During the 16 hour cold soak particles of saturated monoglycerides (SMG’s) and sterol glucosides can be formed which do not dissolve when the temperature is raised back to ambient.
Seta Multi Filtration Tester (MFT) – 91600-4
Features
- Fully automated test
- Interlaboratory study proven with ASTM and EI validated precision
- Large colour touch screen display
- Graphical display of sample flowrate
- Results in less than 15 minutes
- Automatic calculation of FBT
- Storage of test results
- USB and Ethernet connectivity
- RS232 interface for printing
- Simple calibration/verification
specifications
Filter Blocking Tendency (FBT) Range | 1.0 to 30 (low number is best) |
---|---|
Maximum Pressure | 200 kPa |
Power | 110/240 V, 50/60 Hz, 300 W |
Size (HXWXD) | 350 x 300 x 400 mm |
Weight | 9.8 kg |
Manufacturers
show DATA — DO NOT CHANGE ID–
Industries
show DATA — DO NOT CHANGE ID–
Methods
show DATA — DO NOT CHANGE ID–
Spares and accessories for Seta Multi Filtration Tester (MFT) – 91600-4
Start-up Kit, IP387 and D2068 Procedure A – 91615-0
Supplied with: 1 x filter holder, 100 x 1.6 µm filter papers, 1 x tweezers, 2 x anti-splash tubing
Start-up Kit, IP387 and D2068 Procedure B, 98 filters – 91616-0
Supplied with: 98 x 1.6 µm encapsulated glass fibre filters, 6 x female to female, Luer Lock adaptors, Anti-splash tubing
Maintenance Kit – 91645-2
Maintenance kit for use with the Multi Fitration Tester (91600-4). Supplied with: O-rings, Seals, 2 x cleaning wires, Damper/restrictor key
Serial Printer – 80602-0
Serial printer, includes dot matrix printer, cable and roll of paper.
Filter Holder – 91640-0
Filter holder for use with 25 mm filter papers.
Seta Cold Soak test Bath – 91665-0
The Seta cold soak bath is a specially designed bench top unit with programmable digital temperature control suitable for cooling and heating samples in accordance with cold soak test requirements;
Fuel and Receiver Beaker – 91600-002
600 ml glass beaker, one to be used for sample and one to be used for waste with the Seta Multi Filtration Tester (91600-3). Graduated to 400 ml for the purpose of the test method.
Filter IP387 and D2068 Procedure A(Pack of 100) – 91615-001
100 x 1.6 µm filter papers, for D2068 and IP387 Procedure A.
Leur to Leur Adaptor(Pack of 10) – 91616-002
Pack of 10 Leur adaptors for use with the Seta Multi-Filtration Tester (91600-3).
Metal Leur to Leur Adaptor – 90616-003
Leur to leur adaptor for use with filters for ASTM D2068 and IP 387 Procedure B (91615-001).
Certified Encapsulated Filters Procedure B(Pack of 98) – 91616-001
1.6 µm encapsulated filters for performing tests in accordance with IP 387 and D2068 Procedure B. Supplied in a pack of 98 with certificate of verification.
Printer Paper(Pack of 20) – 81002-301
58 mm width. For use with Data Printer (81002-3).
Pressure Calibrator – 91662-0
The digital pressure gauge is designed for hydraulics, pneumatics and mechanical engineering applications. It is quick and easy to install.
Includes calibration certificate.
Measuring Cylinder – 91660-0
500 ml measuring cylinder for verification/calibration.
Filter Blocking Tendency(FBT) Verification Fluid – 91668-0
The Seta Filter Blocking Tendency (FBT) Verification Fluid is used for checking calibration of the Seta Multi-Filtration Tester (91600-3), including verification of filters, according to IP 387 Procedure B and as referenced in IP PM-EA and CEN N403.
The methods recommend use of a certified verification fluid with a nominal FBT = 2.00, a certificate stating the precise value is supplied with the verification material.
The FBT verification material is produced gravimetrically using a balance traceable to internationally recognised standards and is supplied in a 500 ml bottle.
Thermometer Digital: Filtration Tests -199 to 199ºC – 99922-2
Designed specifically for verification of temperatures associated with ASTM D7501 and Cold Soak Filtration tests.