IviCycle C3000
The IviCycle C3000 is designed to be your perfect partner for high channel-count battery applications. The IviCycle offers a low cost-per-channel, high volume testing solution and extensive data storage capability.
The IviCycle is a multi-channel test system with optional impedance capability. The channels are divided over 4 modules. It is possible to mix and match different IviCycle modules to create the desired number of channels with varying compliance.
The C3000 modules each have 4 channels of ±3A/±5V compliance. The modules are available either as a DC version (no EIS), or as an EIS version with impedance capability on each of the channels. This means that simultaneous/parallel impedance measurements are possible on all channels. The IviCycle has an integrated DataSecure module that stores all data for each channel completely independently from the computer. This ensures that if there is a communication loss or a computer crash/update, the measurement will continue and data will not be lost. This provides peace-of-mind for long duration, high throughput testing. The IviCycle unit is built into a 19inch rack mountable housing for easy incorporation into existing storage racks, or as a stand-alone bench-top multi-channel system.
IviCycle C3000
Features
- Multi-channel battery test system
- Each channel: ±3A/±5V
- Optional FRA/EIS: 10µHz-20kHz on each channel for simultaneous testing
- 19″ rack-mountable housing
- Low cost-per-channel battery test solution
Specifications
System performance |
|
---|---|
Current compliance | ±3A |
Maximum applied voltage | ±5V |
Electrode connections |
4; WE, CE, RE, S with 4mm banana plugs
|
Potentiostat bandwidth | >500kHz |
Stability settings | High Speed, Standard and High Stability |
Signal acquisition | Dual channel 18bit ADC, 100,000 samples/s |
Channel combination | No |
Potentiostat |
|
Applied potential range | ±5V, 0.08mV res. |
Applied potential accuracy | 0.2% or 1mV |
Current ranges | ±1mA to ±1A |
Measured current resolution | 0.003% of current range, minimum 0.3pA |
Measured current accuracy | 20pA + 0.025% of FSR |
Galvanostat |
|
Applied current resolution |
0.008% of applied current range
|
Applied current accuracy | 0.2% |
Galvanostatic current ranges | ±1mA to ±10A |
Potential ranges | ±1mV, ±4mV, ±10mV, ±40mV, ±100mV, ±400mV, ±1V, ±4V, ±10V |
Measured potential resolution | 0.0008% of potential range, minimum 7nV |
Measured potential accuracy | 0.2% or 1mV |
Impedance Analyser |
|
Frequency range | 10µHz to 20kHz |
Amplitude | 0.15mV to 1.0V, or 0.03% to 100% of current range |
Electrometer |
|
Input impedance | >1000Gohm //<20pF |
Input bias current | <20pA |
Bandwidth |
>5MHz
|
Peripheral connections |
|
Analog in/out | 2 analog input only, ±10V 16bit resolution, bandwidth 40kHz |
Digital input/output | 1 input (0-5V) / 3 output (0-3.3V) |
Environment |
|
Power requirements | 100-240V, 50-60Hz, 120VA |
Interfacing | USB/LAN (Ethernet) |
Size (w x d x h) | 44 x 26 x 22cm |
Weight | 12.5kg |
PC requirements | Windows 8/10/11, with free USB port |
Integrated DataSecure | Data acquisition time: 2ms minimum, Stored no. of data points: 20M each channel |