HPLC 6000
NWSPEC HPLC 6000 UV-VIS: Cost-Effective, Reliable, and Compliant
This HPLC system offers accurate UV-Vis detection and robust pumping for diverse analytical needs. It’s designed for ease of use with intuitive software, and ensures data integrity with GMP/GLP compliance, including audit trails and e-signatures. Features versatile solvent handling (Isocratic, Binary, Quaternary*) and streamlined workflows.

Features
- Solvent Versatility: Isocratic, Binary & Quaternary* Systems
- Dependable Performance: High-Quality & Reliable
- Streamlined Workflow: Minimize Training, Improved Efficiency
- Accurate Measurements: High Wavelength Accuracy
- Sensitive Detection: High Sensitivity
- Wide Quantification Range: Wide Linear Range
- Enhanced Detection: Unique Pool Structure
- Flexible Wavelength Programming:
Specifications
| LC Liquid Pump | |
| Infusion way | Series dual-plunger pump |
| Flow Rates | 0.001 ~9.999 ml/min |
| Max. Pressure | 6000 psi |
| Pressure Accuracy | ± 2% |
| Flow Accuracy | ≤0.5% |
| Flow Precision | ≤0.2% RSD |
| Pulsation | <15psi from 600 to 1,600 psi, Flow 1mL/min |
| Operating Temperature | 4~40 |
| Weight | 11 kg |
| Dimensions | 1130H x 260W x 420 D, (mm) |
| UV-VIS Spectrophotometric Detector | |
| Wavelength | 190-700 nm |
| Wavelength Accuracy | ≤1 nm |
| Wavelength Reproducibility | ≤± 0.1 nm |
| Spectral bandwidth | 6 nm |
| Flow pool volume | 8 μ L |
| Optical path | 10 mm |
| Dynamic Noise | ≤± 0.75×10- 5 AU (Methanol, 1ml/min, 254nm, 20°C) |
| Static Noise | ≤± 0.5×10- 5 AU(Empty pool, Response time 1 s, 20 °C) |
| Dynamic Drifting | ≤± 1×10- 4 AU (Methanol, 1ml/min, 254nm, 20°C) |
| Static Drifting | ≤± 0.5×10- 4 AU(Empty pool, Response time 1 s, 20 °C) |
| Linear range | ≥104 |
| Minimum detectable concentration | ≤4×10-⁹ g/mL (Naphthalene/Methanol solution) |
| Qualitative repeatability | RSD ≤ 0.1% |
| Weight | 11 kg |
| Dimensions | 130H x 260W x 420 D (mm) |