Ivium-n-Stat
The instrument is designed for a wide range of applications, such as battery/fuel cell-testing, electrode development, biotechnology and other Electrochemical Research applications. The Ivium-n-Stat offers a complete package. The hardware includes a built-in high-performance Frequency Response Analyser, low current measurement possibility down to 100pA full-scale, and all the standard electrochemical techniques. All measurement and dataprocessing software is included (for Windows based PCs). The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented. Multiple analog and digital input & output ports are available for monitoring & control of peripheral equipment. This functionality is fully integrated in the software. Each channel can be expanded with various options and modules (depending on the compatibility), such as a power booster, multiplexer, etc.
- Data from different channels plotted individually or in 1 single graph
Ivium-n-Stat
Features
- Up to 8 modules/16 channels per frame (stackable up to 64 channels) can be installed
- Modules can be mix and matched for maximum versatility
- Multi-channel Potentiostat/Galvanostat/ZRA
- Integrated impedance analyser in each channel as standard: 10µHz – 250kHz (1MHz optional)
Specifications
Environment |
|||||
PC requirements |
Windows 8/10/11, with free USB port
|
||||
Available modules |
|||||
sModule 2.5A/10V |
sModule 2.5A/10V | sModule 2.5A/10V | sModule 2.5A/10V |
sModule 2.5A/10V |
|
System performance |
|||||
Number of channels in module |
1 |
1 | 1 | 1 |
2 |
Current compliance |
±2.5A |
±5A | ±10A | ±20A |
±2.5A |
Maximum compliance voltage |
±10V |
±10V | ±5V | ±2V |
±10V |
Maximum applied voltage |
±10V |
±10V | ±5V | ±2V |
±10V |
Potentiostat |
|||||
Applied potential range |
±10V, 0.08mV res. |
±10V, 0.08mV res. | ±5V, 0.08mV res. | ±2V, 0.08mV res. |
±10V, 0.08mV res. |
Current ranges |
±100pA to ±10A |
±100pA to ±10A | ±100pA to ±10A | ±100pA to ±10A |
±100pA to ±10A |
Galvanostat |
|||||
Galvanostatic current ranges |
±10nA to ±10A |
±10nA to ±10A | ±10nA to ±10A | ±10nA to ±10A |
±10nA to ±1A |
Peripheral connections |
|||||
Analog in/out |
2/1 (±10V, 16bit resolution, bandwidth 40kHz) |
2/1 (±10V, 16bit resolution, bandwidth 40kHz) | 2/1 (±10V, 16bit resolution, bandwidth 40kHz) | – |
±10V input only, 100GOhm, 16bit resolution, bandwidth 40kHz |
Digital input/output |
1 input (0-5V) / 3 output (0-3.3V) |
1 input (0-5V) / 3 output (0-3.3V) | 1 input (0-5V) / 3 output (0-3.3V) | – |
– |
I-out/E-out |
Analog monitor for cell current and potential |
Analog monitor for cell current and potential | Analog monitor for cell current and potential | – |
– |
AC-out |
±0.5V sinewave 10µHz-250kHz with variable attenuation |
±0.5V sinewave 10µHz-250kHz with variable attenuation | ±0.5V sinewave 10µHz-250kHz with variable attenuation | – |
– |
Channel-X input, Channel-Y input |
±4V: to record impedance from peripheral devices |
±4V: to record impedance from peripheral devices | ±4V: to record impedance from peripheral devices | – |
– |
Peripheral port |
Yes |
Yes | Yes | – |
– |
Manufacturers