PRODUCT

  1. Home
  2. chevron_right
  3. Special Offers
  4. chevron_right
  5. HPLC 6000

HPLC 6000

NWSPEC HPLC 6000 UV-VIS: Cost-Effective, Reliable, and Compliant

This HPLC system offers accurate UV-Vis detection and robust pumping for diverse analytical needs. It’s designed for ease of use with intuitive software, and ensures data integrity with GMP/GLP compliance, including audit trails and e-signatures. Features versatile solvent handling (Isocratic, Binary, Quaternary*) and streamlined workflows.

Do NOT CHANGE OR REMOVE THIS ELEMENT

Features

  • Solvent Versatility: Isocratic, Binary & Quaternary* Systems
  • Dependable Performance: High-Quality & Reliable
  • Streamlined Workflow: Minimize Training, Improved Efficiency
  • Accurate Measurements: High Wavelength Accuracy
  • Sensitive Detection: High Sensitivity
  • Wide Quantification Range: Wide Linear Range
  • Enhanced Detection: Unique Pool Structure
  • Flexible Wavelength Programming:

Specifications

LC Liquid Pump
Infusion way Series dual-plunger pump
Flow Rates 0.001 ~9.999 ml/min
Max. Pressure 6000 psi
Pressure Accuracy ± 2%
Flow Accuracy ≤0.5%
Flow Precision ≤0.2% RSD
Pulsation <15psi from 600 to 1,600 psi, Flow 1mL/min
Operating Temperature 4~40
Weight 11 kg
Dimensions 1130H x 260W x 420 D, (mm)
UV-VIS Spectrophotometric Detector
Wavelength 190-700 nm
Wavelength Accuracy ≤1 nm
Wavelength Reproducibility ≤± 0.1 nm
Spectral bandwidth 6 nm
Flow pool volume 8 μ L
Optical path 10 mm
Dynamic Noise ≤± 0.75×10- 5 AU (Methanol, 1ml/min, 254nm, 20°C)
Static Noise ≤± 0.5×10- 5 AU(Empty pool, Response time 1 s, 20 °C)
Dynamic Drifting ≤± 1×10- 4 AU (Methanol, 1ml/min, 254nm, 20°C)
Static Drifting ≤± 0.5×10- 4 AU(Empty pool, Response time 1 s, 20 °C)
Linear range ≥104
Minimum detectable concentration ≤4×10-⁹ g/mL (Naphthalene/Methanol solution)
Qualitative repeatability RSD ≤ 0.1%
Weight 11 kg
Dimensions 130H x 260W x 420 D (mm)

Manufacturers

show DATA — DO NOT CHANGE ID–

Industries

show DATA — DO NOT CHANGE ID–